|
The certification process involves an evaluation of a number of facets
of each laboratory seeking certification. In addition to specific
criteria and/or general guidelines given in the laboratory certification
manual and analytical procedures the following quality assurance checks
are required or recommended. A detailed written explanation or
authorized substitution may be required for recommendations which are
not followed.
BOD5 (CBOD5)
DO meters require calibration each day of use. (Required)
When "air calibrating" the DO meter use the true barometric pressure at
the time of calibration instead of the average barometric pressure based
on altitude. (Recommended)
Dilution water blanks are necessary with each set of BOD (CBOD)
analysis. (Required) If nitrification inhibitor is used in samples it
must also be added to at least one distilled water blank. (Recommended)
Periodic check of dilution water quality, seed effectiveness, and
analytical technique are required by making a BOD measurement on pure
organic compounds (normally a glucose/glutamic acid mixture).
(Required) How often this is done depends upon the sample load of the
laboratory, but in general if a laboratory analyzes samples once a week
or less frequently, it should be done at least quarterly. If the sample
load is such that the laboratory analyzes samples daily, the
glucose/glutamic acid check should be done at least weekly.
(Recommended)
Duplicate analysis should be run on at least one sample each time BODs
are analyzed. If more than ten samples are being analyzed one of each
ten samples should be set up with a duplicate set of dilutions.
(Recommended)
Ammonia (NH3-N) as N
EPA requires samples be distilled prior to analysis unless comparative
data for each sample source is on file at the laboratory demonstrating
comparable analytic results between distilled and non-distilled portions
of the sample. (Required) Comparative data should be based on a
sufficient number (generally 30 required, but may be less in some
circumstances) analytical comparisons performed over a sufficient time
period to validate NH3 levels as they change with seasons. After
initial comparison is completed periodic comparisons should be performed
to insure comparability has not changed. (Recommended)
A reagent blank, a check (reference) standard, duplicate and spiked
quality control samples should be run each time samples are analyzed.
(Recommended)
Total Suspended Solids (TSS)
If using EPA or SM-18 method, filters weight must be shown to be at
constant weight both after washing and after filtration and drying of
sample. Constant weight is defined as weight change of 4% or 0.5 mg
(0.0005g) whichever is LESS. (Required)
Quality control should consist of duplicate analysis of 10 % of samples
analyzed and quarterly analysis of a known concentration reference
sample. (Recommended)
|